Anesthetic Equipment and Monitoring

Anesthetic Equipment and Monitoring

The Operating Room Environment

Anaesthetists spend much of their working day in the operating theatre and share responsibility for protecting patients and staff from its hazards, several of which are unique to this environment. The anaesthetist must therefore understand the medical gas supply, fire prevention, electrical safety and environmental factors such as temperature, humidity and ventilation. This lesson covers the theatre features and hazards of special interest to anaesthetists, then the breathing systems, the anaesthetic workstation and monitoring.

Culture of Safety

The theatre feels safe, but errors occur unless the whole team stays vigilant. Harm is best prevented by a culture of safety in which unsafe acts are spotted and stopped before they reach the patient. A central tool is the surgical safety checklist, used for every patient before induction, before incision and before the patient leaves theatre.

  • The WHO Surgical Safety Checklist (2008) has three phases: Sign In (before induction), Time Out (before skin incision) and Sign Out (before the patient leaves theatre). In the UK it sits within local and national safety standards for invasive procedures (LocSSIPs and NatSSIPs), which add a team brief before the list and a debrief after it.
  • Checklists work best when used interactively, with the whole team stopping and paying attention.
  • Rather than reading the whole list and asking for general agreement, the leader should ask for a specific answer to each item (e.g., "Can everyone confirm this patient is John Doe, for a left inguinal hernia repair?").
  • A good checklist covers the key points only and can be completed quickly (e.g., in under 90 seconds).
  • Correctly used, checklists reduce preventable complications such as wrong-site surgery, retained foreign objects and drug errors in patients with known allergies. Anaesthetists have long led in patient safety and should champion their use.

Medical Gas Systems

The medical gases piped to theatre are oxygen, nitrous oxide and medical air; in the UK there is also surgical air at a higher pressure for power tools, while some countries pipe nitrogen for the same purpose. Vacuum for surgical suction and the anaesthetic gas scavenging system (AGSS; called waste anaesthetic gas disposal, WAGD, in the US) complete the system. The anaesthetist must understand where these gases come from and how they reach the machine in order to prevent or detect depletion, contamination or misconnection; failure of the oxygen supply in particular can kill. System design follows HTM 02-01 in the UK (formerly HTM 2022), European standards (EN ISO 7396-1) and NFPA 99 in the United States.

SOURCES OF MEDICAL GASES

Oxygen

Medical oxygen (at least 99.5% pure) is made by fractional distillation of liquefied air. It is stored either as a compressed gas in cylinders or as a refrigerated liquid.

  • Cylinder manifolds: Smaller hospitals use two banks of large cylinders joined by a manifold. One bank is in use at a time; the manifold regulates the cylinder pressure (about 13,700 kPa / 137 bar (2000 psig); newer cylinders are filled to 23,000 kPa / 230 bar) down to pipeline pressure (about 400 kPa (4 bar) in the UK and Europe; 345–380 kPa (50–55 psig) in the US) and switches automatically to the reserve bank when the first is exhausted.
  • Liquid oxygen (vacuum-insulated evaporator, VIE): Larger hospitals store oxygen as a liquid, which is cheaper per litre. A gas can be liquefied by pressure only below its critical temperature, which for oxygen is −119 °C; in the VIE the liquid is kept at about −160 °C and 5–10 bar. One litre of liquid yields about 840 L of gas. A reserve (cylinder manifold or a second vessel) holds enough for at least a day.
  • Oxygen concentrators: Used in some hospitals, remote sites and homes. Zeolite adsorbs nitrogen under pressure (pressure-swing adsorption), giving about 90–96% oxygen; the remainder is mainly argon.
  • Emergency supply: An emergency oxygen cylinder must always be available on or beside the anaesthetic machine. Oxygen is a gas in the cylinder, so the pressure falls in proportion to the contents.
  • Cylinder safety: Cylinders on the machine are pin-indexed so that only the correct gas fits the yoke. Cylinders carry pressure-relief devices: a fusible plug of Wood's metal (a low-melting alloy) that melts if a fire heats the cylinder, and/or a frangible (bursting) disc that ruptures at a set pressure (about 22,800 kPa (3300 psig) in US E-cylinders), well below the pressure the cylinder wall withstands. Cylinders are hydraulically tested, and the test date, tare weight and contents are marked on them.

Nitrous Oxide

Nitrous oxide is stored in cylinders, usually on a manifold with automatic changeover. Its critical temperature (36.5 °C) is above room temperature, so it is kept as a liquid under pressure without refrigeration; machine cylinders also contain liquid with vapour above it.

  • While liquid remains and the temperature is constant, the gauge shows the saturated vapour pressure: about 5200 kPa / 52 bar (745 psig) at 20 °C (about 44 bar at 15 °C). The pressure falls only when the liquid is used up, by which time roughly a quarter of the contents (about 400 L in an E-cylinder of about 1590 L) remains.
  • Cylinders are filled by weight to a filling ratio (mass of gas ÷ mass of water the cylinder could hold) of 0.75 in temperate climates and 0.67 in hot climates, leaving space for the liquid to expand. A reading clearly above the expected value suggests a faulty gauge, overfilling or the wrong gas.
  • Evaporation of the liquid takes energy (latent heat of vaporisation), so the cylinder cools. Vapour pressure and cylinder pressure fall, and at high flows frost forms on the cylinder and the regulator can freeze.
  • Like oxygen cylinders, nitrous oxide cylinders carry pressure-relief devices.

DELIVERY OF MEDICAL GASES

Medical gases travel from the central supply to theatre through seamless, degreased copper pipes. The network is designed so that the pressure drop across it stays small (no more than about 35 kPa (5 psig) at design flow). In theatre the outlets appear as wall-mounted terminal units, ceiling hose drops, pendants or articulated arms.

Pipeline Connections

  • Terminal units (wall outlets): Each gas has a self-sealing socket that accepts only…

Environmental Factors in the Operating Room

TEMPERATURE

Operating theatres often feel cold to awake patients and anaesthetists, while surgeons and scrub staff stand under hot lights in gowns for long periods. Staff comfort must be balanced against the patient's needs. UK theatre design guidance (HTM 03-01) allows the temperature to be set between about 18 °C and 25 °C; US guidance gives 20–24 °C (68–75 °F) for adults. The effect of the room on the patient's core temperature must be watched, because inadvertent perioperative hypothermia (core temperature below 36.0 °C) increases wound infection, coagulopathy and blood loss, cardiac events, shivering and oxygen consumption, drug duration and length of stay.

  • NICE guidance (CG65) for…

Electrical Safety

The extensive use of electrical equipment in theatre exposes patients and staff to electric shock, burns and fire. Anaesthetists must understand the hazards and the protective systems built into equipment and theatre wiring.

THE RISK OF ELECTROCUTION

An electric shock occurs when the body completes a circuit between two conductors at different potentials. Usually one contact is a live conductor (mains is 230 V at 50 Hz in the UK and Europe, 120 V at 60 Hz in North America) and the circuit is completed through earth (ground). Mains supplies are earth-referenced: the neutral is connected to earth at the substation, so a person who is earthed (e.g., touching a metal part of the operating table) needs…

SURGICAL DIATHERMY (ELECTROCAUTERY, ELECTROSURGERY)

Surgical diathermy units (electrosurgical units, ESUs; "electrocautery") are the main deliberate source of electrical current through the patient in theatre. In monopolar diathermy a radiofrequency current passes from a small active electrode (the diathermy tip or forceps) through the patient and leaves through a large plate, the patient return electrode (dispersive pad, "diathermy plate"), back to the generator.

Mechanism of Action and Safety

  • Frequency and fibrillation: Diathermy uses…

Surgical Fires & Thermal Injury

FIRE PREVENTION & PREPARATION

Surgical fires are rare but almost always preventable. Unlike most complications they follow simple physics and chemistry: when the right combination of factors is present a fire will start, and removing any one factor prevents it.

Every fire needs the three elements of the fire triad (fire triangle):

  • Fuel: alcohol-based skin preparations, drapes, gowns, swabs and gauze, tracheal tubes and airway devices, oxygen masks and cannulae, ointments, the patient's hair, and bowel gas.
  • Oxidiser: oxygen and nitrous oxide (nitrous oxide breaks down with heat to release oxygen, so it…

CREW RESOURCE MANAGEMENT: CREATING A CULTURE OF SAFETY

Crew resource management (CRM) began in aviation to improve teamwork and reduce human error. Its aim is to empower every team member to question or stop anything they believe is unsafe. In theatre this means that anyone, whatever their seniority, may raise a safety concern and see it resolved before a mistake reaches the patient. CRM is usually described by seven principles. In anaesthesia the same skills are…

ROLES OF ACCREDITATION AGENCIES & REGULATORY BODIES

Hospitals are inspected and regulated so that patients receive care that meets agreed standards. The bodies differ by country, but the principles are the same: an external organisation checks whether a hospital has appropriate policies and whether they are followed, usually through a self-assessment followed by a site visit in which inspectors observe practice, examine records and compare what they find with standards and best practice.

UK and European Bodies

  • Care Quality Commission (CQC): the…

Breathing Systems

Breathing systems are the final link between the anaesthetic machine and the patient, delivering oxygen and anaesthetic gases and removing carbon dioxide. Many designs exist, differing in efficiency, convenience and complexity. This part of the lesson covers insufflation, open-drop and draw-over systems, the Mapleson systems, the circle system and resuscitation systems. The traditional labels open, semi-open, semi-closed and closed are used inconsistently and are best…

MAPLESON CIRCUITS

Insufflation and draw-over systems have several drawbacks: poor control of the inspired concentration (and so of anaesthetic depth), awkward equipment during head and neck surgery, and heavy pollution of the theatre with waste gas. The Mapleson systems overcame some of these problems by adding breathing tubes, a fresh gas inlet, an adjustable pressure-limiting (APL) valve and a reservoir bag. The position of these components relative to the patient determines how each system…

Performance Characteristics of Mapleson Circuits

Mapleson systems are simple, light and cheap. Their efficiency is judged by the fresh gas flow (FGF) needed to prevent rebreathing of carbon dioxide. They have no unidirectional valves and no absorber, so rebreathing is prevented only by a fresh gas flow large enough to flush exhaled gas (containing carbon dioxide) out through the APL valve before the next breath. Some rebreathing usually occurs, and the total FGF determines how much. Flows are therefore high, which wastes vapour, cools and dries the airway, and needs good…

THE CIRCLE SYSTEM

Mapleson systems solve some problems of insufflation and draw-over, but they need high fresh gas flows to prevent rebreathing of carbon dioxide. The high flow wastes anaesthetic agent, pollutes the theatre and causes large losses of heat and moisture from the patient. The circle system overcomes these problems by adding a carbon dioxide absorber, which allows exhaled gas to be rebreathed safely once the carbon dioxide has been removed.

Components of the Circle System

The circle system consists of:

  • A carbon dioxide absorber…

Optimization and Performance of the Circle System

Optimisation of Circle System Design

The arrangement of the components determines how efficient and safe a circle is. The preferred modern arrangement is:

  • Unidirectional valves: placed relatively close to the patient to limit backflow into the inspiratory limb if a breathing tube leaks, but not in the Y-piece, where their function cannot easily be seen and checked during anaesthesia (and where they would add weight at the patient end).
  • Fresh gas inlet: between the absorber and the inspiratory valve. Downstream of the inspiratory valve, fresh gas would bypass the patient and be vented during expiration.…

RESUSCITATION BREATHING SYSTEMS

Resuscitation bags (self-inflating bags, bag-valve-mask units, often called "Ambu bags") are widely used for emergency ventilation because they are simple and portable, work without any gas supply, and with an oxygen reservoir can deliver an oxygen concentration close to 100%. They differ fundamentally from Mapleson and circle systems because they contain a non-reb…

The Anesthesia Workstation

No piece of equipment is more closely linked to anaesthesia than the anaesthetic machine. The anaesthetist uses it to control ventilation, deliver precise oxygen concentrations and give inhalational agents, so its correct function is vital to patient safety. Modern machines, called anaesthetic workstations, include many built-in safety features, monitors and microprocessors, and their modular design allows many configurations, so anaesthetists must know…

FLOW CONTROL CIRCUITS

Pressure Regulators

Cylinder pressure is high and falls as the cylinder empties (e.g., from about 13,700 kPa (2000 psig) in a full oxygen cylinder), which would make flow control difficult and dangerous. A pressure regulator (reducing valve) on each cylinder inlet therefore reduces it to a constant lower pressure. A regulator works by balancing the force of a spring on a diaphragm against the gas pressure on the other side, which opens or closes a valve between the high- and low-pressure chambers.

  • The regulated cylinder pressure is set slightly below pipeline pressure (in American machines 45–47 psig, about 310–325 kPa, against a pipeline of 50 psig, about 345 kPa; European machines work on the same principle with a pipeline of about 400…

Vaporizers

Volatile anaesthetics (sevoflurane, desflurane, isoflurane, halothane) must be converted from liquid to vapour before they can be given. Vaporisers are concentration-calibrated devices that add an accurately known concentration of agent to the fresh gas flow. On the anaesthetic machine they sit in the low-pressure circuit, between the flowmeters and the common gas outlet (vaporiser outside the circle). Unless the machine takes only one vaporiser, an interlock or exclusion device (e.g., the Selectatec back bar system) must prevent more than one vaporiser being turned on at once, which could otherwise give a dangerous overdose of two agents.

A. Physics of Vaporisation

In a closed container, molecules of a volatile liquid distribute between liquid and vapour phases until equilibrium. Vapour molecules striking the walls exert the saturated vapour pressure (SVP). The SVP does not depend on the volume of the container or on the ambient pressure; it depends only…

Common (Fresh) Gas Outlet

Whereas the machine has several gas inlets, it has only one common gas outlet (fresh gas outlet), which delivers the final mixture of gases and vapour from the flowmeters and vaporisers to the breathing system. It is a standard conical fitting (22 mm male, 15 mm female) and usually has…

THE BREATHING CIRCUIT

In adults the breathing system used with the anaesthetic machine is usually the circle. The gas composition at the common gas outlet, which is precisely controlled, can differ greatly from that inside the breathing system. The composition in the circle depends also on the patient's uptake of agent and oxygen, minute ventilation, total fresh gas flow, any leaks and the internal volume of the system. High flows during induction and emergence reduce these differences. Measuring both inspired and end-tidal oxygen, carbon dioxide and agent concentrations is essential for safe anaesthesia, and the Association of Anaesthetists monitoring standards (2021) require continuous monitoring of all three during general anaesthesia, together with airway pressure and volume.

Oxygen Analysers

General…

VENTILATORS

All modern anaesthetic workstations include a mechanical ventilator. Theatre ventilators used to be much simpler than intensive care ventilators, but the difference has narrowed as technology has advanced and sicker patients need intensive care-level ventilation in theatre. Ventilators produce gas flow by creating a pressure gradient between the proximal airway and the alveoli. Their function is best understood by the four phases of the ventilatory cycle.

Phases of the Ventilatory Cycle

A. Inspiratory Phase

During inspiration the ventilator delivers the tidal volume. Ventilators are classified by how they behave in inspiration:

  • Pressure generators produce a set pressure waveform…

Pressure & Volume Monitoring

Monitoring airway pressure and volume is essential for judging lung mechanics and keeping ventilation safe. The shape of the breathing system pressure waveform gives important diagnostic information, and modern workstations display pressure, flow and volume against time and as pressure–volume and flow–volume loops.

Peak vs. Plateau Pressure

Two pressures are routinely assessed during volume-controlled ventilation with…

Problems Associated with Anesthesia Ventilators

A. Ventilator–Fresh Gas Flow Coupling

In a traditional double-circuit (bellows) ventilator the ventilator spill (exhaust) valve is closed during inspiration and the APL valve is cut out of the circuit. Any fresh gas flow (FGF) arriving from the common gas outlet during inspiration therefore has nowhere to go except into the patient, and it adds to the set tidal volume. This is called coupling.

Worked example: FGF 6 L/min (6000 mL/min), rate 10 breaths/min, I:E ratio 1:2 (inspiration is one third of each cycle):

  • Fresh gas per cycle: $6000 \div 10 = 600$ mL


  • Fresh gas entering during inspiration…

WASTE-GAS SCAVENGERS

Waste-gas scavengers remove gases vented from the breathing system, mainly through the APL valve (spontaneous or manual ventilation) and the ventilator spill valve (mechanical ventilation). Sidestream gas analysers also discharge sampled gas, which must be returned to the circuit or scavenged. Without scavenging, anaesthetic gases pollute the operating theatre and expose staff.

Occupational exposure limits are national, not international:

…

ANESTHESIA MACHINE CHECKOUT LIST

Misuse or malfunction of anaesthetic gas delivery equipment can cause major morbidity and death. Checking the equipment before use confirms that it works and makes the anaesthetist familiar with it. Checks must follow a national guideline and be adapted to the specific machine and the manufacturer's instructions.

  • UK: the Association of Anaesthetists guideline Checking Anaesthetic Equipment (2019) — a full check at the start of every session and a shorter check before each case, with a record in a log book or the machine's electronic log.
  • Europe: national society checklists based on the same principles; ESAIC supports structured pre-use checks as part of the Helsinki Declaration on Patient Safety in Anaesthesiology.
  • USA: a generic FDA checkout (1993), later replaced by the 2008 ASA…

Cardiovascular Monitoring

ARTERIAL BLOOD PRESSURE

Rhythmic ejection of blood from the left ventricle into the arterial tree produces a pulsatile arterial pressure.

  • Systolic pressure (SBP): the peak pressure during ventricular ejection.
  • Diastolic pressure (DBP): the lowest pressure during diastole.
  • Pulse pressure: SBP − DBP; it rises with stroke volume and arterial stiffness.
  • Mean arterial pressure (MAP): the time-weighted average pressure over one cardiac cycle; it is the driving pressure for organ perfusion.…

Continuous Noninvasive Arterial Blood Pressure

Besides intermittent cuff methods, two techniques give continuous, beat-to-beat non-invasive blood pressure.

Arterial Tonometry

Arterial tonometry measures beat-to-beat pressure from the force needed to partially flatten (applanate) a superficial artery lying on bone, typically the radial artery at the wrist. A sensor containing an array of pressure transducers is placed over the artery; the element directly over the vessel senses the intraluminal pressure as…

Invasive Arterial Blood Pressure

A. Selection of Artery for Cannulation

Several peripheral arteries can be cannulated percutaneously; the choice depends on the patient, the surgery and the collateral circulation.

  • 1. Radial artery: the most commonly used site: superficial, easy to compress, and with good collateral flow because the ulnar artery (usually larger) supplies the hand through the palmar arches in most people; a small proportion have incomplete arches. The Allen test (the patient clenches the fist, both arteries are…

Clinical Considerations for Invasive Arterial Monitoring

Intra-arterial cannulation is the reference technique for continuous beat-to-beat blood pressure, but its accuracy depends on the dynamic characteristics of the whole measuring system (cannula, tubing, stopcocks, flush device and transducer). A poorly performing system gives false readings that can lead to wrong and dangerous treatment.

System Dynamics: Natural Frequency and Damping

By Fourier analysis, the arterial waveform is the sum of a fundamental sine wave (at the heart rate) and its harmonics. Faithful…

ELECTROCARDIOGRAPHY

Indications and Contraindications

Continuous electrocardiogram (ECG) monitoring is mandatory for every patient during anaesthesia: it is part of the Association of Anaesthetists and ESAIC minimum monitoring standards (and the ASA standards in the USA), from before induction until recovery. There are no contraindications.

Techniques and Complications

Diagnostic sensitivity depends on lead selection. Electrodes on the chest and…

CENTRAL VENOUS PRESSURE

Indications

Central venous catheterisation and central venous pressure (CVP) monitoring are indicated for:

  • Monitoring of right heart filling pressure and its trend, and sampling of central venous oxygen saturation ($ScvO_2$).
  • Administration of drugs that need rapid dilution in a large vein: vasopressors and inotropes, chemotherapy, hyperosmolar solutions, parenteral nutrition, concentrated potassium.
  • Aspiration of entrained air in procedures at high risk of venous air embolism (sitting-position neurosurgery), using a multi-orifice catheter with its tip at the junction of the superior vena cava and right atrium.
  • Access for pulmonary artery catheters and transvenous pacing wires.
  • Poor peripheral venous access…

PULMONARY ARTERY CATHETERIZATION

Indications

The flow-directed pulmonary artery (PA) catheter (Swan–Ganz) was introduced in 1970 and became central to haemodynamic monitoring. Its main advantage was measurement of the pulmonary capillary occlusion ("wedge") pressure (PCOP, PAOP or PCWP). With the balloon inflated, a static column of blood connects the catheter tip to the left atrium, so PCOP estimates left atrial pressure and, by extension, left ventricular end-diastolic pressure (LVEDP), used as a surrogate for left ventricular preload. Normal PCOP is about 0.8–1.6 kPa (6–12 mmHg).

Combined with cardiac output (CO) from thermodilution, it allows calculation of stroke volume…

CARDIAC OUTPUT

Indications

Measurement of cardiac output (CO), allowing calculation of stroke volume (SV) and systemic vascular resistance (SVR), was one of the main reasons for PA catheterisation. Many less invasive methods now estimate ventricular performance and guide goal-directed fluid and vasoactive therapy. Normal resting values in an adult: CO 4–8 L/min, cardiac index (CO/body surface area) 2.5–4 L/min/m², SV 60–100 mL.

Techniques and Complications

A. Thermodilution

The reference clinical method with a PA catheter. A known volume (5 or 10 mL; 2.5 mL in children) of cold fluid (iced or room-temperature 5% glucose or 0.9% saline) is injected rapidly into the right atrium through the proximal (CVP) port. It mixes with blood and causes a transient fall in temperature, sensed by the thermistor near the tip…

Echocardiography

Transthoracic (TTE) and transoesophageal echocardiography (TOE) are the most powerful tools for perioperative diagnosis and assessment of cardiac function. TOE is especially suited to the operating theatre because it gives continuous, detailed images of the heart when access to the chest is limited.

Main perioperative uses:

  • Diagnosis of haemodynamic instability (systolic or diastolic ventricular failure, hypovolaemia, pericardial tamponade, left ventricular outflow tract obstruction, pulmonary embolism).
  • Detection of myocardial ischaemia as new regional wall motion abnormalities (RWMAs), which appear earlier and more…

Other Monitoring Methods

This part of the lesson covers the non-cardiovascular monitors used perioperatively: respiratory gas exchange, neurological function, neuromuscular transmission and body temperature. They are essential for patient safety and for detecting adverse events early.

Respiratory Gas Exchange Monitors

Under the Association of…

PULSE OXIMETRY

Indications and Contraindications

Pulse oximetry is a mandatory monitor for every anaesthetic, including sedation and regional anaesthesia, and from before induction through recovery (Association of Anaesthetists, ESAIC and WHO standards; part of the WHO Surgical Safety Checklist). There are no contraindications.

Techniques and Complications

A pulse oximeter estimates arterial haemoglobin oxygen saturation non-invasively and continuously ($SpO_2$) by combining spectrophotometric oximetry with plethysmography. The probe holds two (or more) light-emitting diodes (LEDs) and one photodiode detector, placed across a perfused tissue that light can pass through (finger, toe, earlobe, nose…

CAPNOGRAPHY

Indications and Contraindications

Continuous measurement of end-tidal $CO_2$ ($ETCO_2$) with a waveform is mandatory during every general anaesthetic, for every patient with a tracheal tube or supraglottic airway wherever they are (theatre, intensive care, emergency department, transfer), and during deep sedation (Association of Anaesthetists, ESAIC, DAS and ERC guidance; the NAP4 audit linked the absence of capnography to many airway deaths in intensive care). There are no contraindications.

Capnography is a monitor of the lungs, the circulation and the breathing system:

  • It is the most rapid and reliable way to confirm tracheal (or…

ANESTHETIC GAS ANALYSIS

Indications

Continuous analysis of inspired and expired anaesthetic agent concentrations is essential whenever an inhalational agent is used (Association of Anaesthetists and ESAIC minimum monitoring), to confirm delivery, avoid overdose and awareness, and allow safe low-flow anaesthesia — at low fresh gas flows the circle concentration differs greatly from the vaporizer setting. Most monitors also display the end-tidal value as a multiple of age-adjusted MAC. There are no contraindications. Agent…

Neurological System Monitors

ELECTROENCEPHALOGRAPHY

Indications and Contraindications

The electroencephalogram (EEG) records the summed postsynaptic potentials of pyramidal cells in the cerebral cortex. Raw or processed EEG is used to monitor brain function and anaesthetic depth. Indications include:

  • Carotid endarterectomy and other cerebrovascular surgery under general anaesthesia: to detect cerebral ischaemia during cross-clamping and guide shunt insertion.
  • Cardiac and aortic surgery with deep hypothermic circulatory arrest: to confirm burst suppression or an isoelectric EEG (maximal suppression of cerebral metabolism) before arrest.
  • Neurosurgery and…

Neurological System Monitors (Continued)

EVOKED POTENTIALS

Indications

Intraoperative evoked potential (EP) monitoring is indicated for surgery with a high risk of neurological injury: scoliosis and other spinal surgery, spinal cord tumours, thoracoabdominal aortic aneurysm repair, posterior fossa and brainstem surgery (acoustic neuroma), carotid surgery and some intracranial aneurysms. It assesses the functional integrity of sensory or motor pathways by recording the electrophysiological response to stimulation, so that injury is detected early enough to intervene. It replaces or supplements the intraoperative wake-up test.

Techniques and Complications

A neural pathway is stimulated repeatedly and the…

Temperature Monitoring

Indications

Temperature should be monitored during all but the shortest anaesthetics. UK standards (Association of Anaesthetists; NICE guideline on inadvertent perioperative hypothermia) require measurement before induction and at least every 30 minutes during surgery, and temperature monitoring for procedures lasting more than 30 minutes; continuous core monitoring is best. Inadvertent perioperative hypothermia is a core temperature below 36.0 °C. Postoperative temperature is used as a marker of anaesthetic quality.

  • Hypothermia: delayed drug metabolism (prolonged neuromuscular…

URINARY OUTPUT

Indications

Urinary bladder catheterisation (with a Foley catheter and an hourly urometer) is the only reliable way to measure urine output during surgery. It is routine for many long and complex procedures:

  • Cardiac surgery
  • Aortic or renal vascular surgery
  • Craniotomy, especially when mannitol or other diuretics are given
  • Major abdominal and pelvic surgery
  • Procedures with large…

PERIPHERAL NERVE STIMULATION

Indications

Sensitivity to neuromuscular blocking drugs varies widely between patients. Neuromuscular function should therefore be monitored in every patient given a neuromuscular blocking drug, from induction until recovery: the Association of Anaesthetists (2021) and ESAIC (2023) recommend quantitative monitoring whenever these drugs are used. Nerve stimulation also shows the onset of block before intubation, guides top-up doses and the choice and dose of reversal, and confirms recovery before extubation.